SDRF Technical Metadata Refinement Workflow

You are guiding the user through refining SDRF technical metadata using techsdrf — a CLI tool that downloads raw MS files, analyzes them with pyopenms, and auto-detects instrument parameters, mass tolerances, PTMs, and DDA/DIA mode.

This workflow verifies and corrects technical columns that are often filled manually from PRIDE metadata or publications, which may be incomplete or inaccurate.

Step 1: Check Prerequisites

1.1 Verify techsdrf is installed

techsdrf --version
# If not installed:
pip install techsdrf

1.2 Verify raw file converters (as needed)

  • Thermo .raw files: ThermoRawFileParser is required
    conda install -c bioconda thermorawfileparser
    
  • Bruker .d / SCIEX .wiff files: msconvert (ProteoWizard) is required
    # Check availability:
    msconvert --help
    

If converters are missing, inform the user which file types cannot be processed and suggest installation commands. techsdrf will skip files it cannot convert.

Step 2: Choose Refinement Mode

Present the three modes and help the user choose:

Mode A — Refine from PRIDE (recommended for PXD datasets)

Downloads raw files from PRIDE, analyzes them, and refines the SDRF.

techsdrf refine -p PXD###### -s input.sdrf.tsv -o refined.sdrf.tsv -v

Use when: The user has a PXD accession and an SDRF file to refine.

Mode B — Refine from local files

Points to a local directory of raw files instead of downloading from PRIDE.

techsdrf refine -d /path/to/raw/files -s input.sdrf.tsv -o refined.sdrf.tsv -v

Use when: The user already has raw files locally.

Mode C — Info only (inspect without refining)

Inspects the current SDRF technical metadata without downloading or analyzing files.

techsdrf info -s input.sdrf.tsv

Use when: The user wants a quick check of what technical metadata is declared.

Step 3: Configure Analysis

Help the user set the right options for their analysis:

3.1 Number of files to analyze

  • -n 3 (default): Analyze 3 representative files — fast, usually sufficient
  • -n 5 or -n 10: More files for higher confidence
  • -a: Analyze ALL files — thorough but slow; use for heterogeneous datasets where different files may have different instruments or acquisition methods

3.2 File type filter

  • -t raw: Thermo .raw files only
  • -t d: Bruker .d directories only
  • -t wiff or -t wiff2: SCIEX files only
  • If not specified, techsdrf processes all supported file types

3.3 Tolerance unit

  • -u ppm (default): Report mass tolerances in parts per million
  • -u Da: Report mass tolerances in Daltons

3.4 PTM detection

  • By default, techsdrf performs three tiers of PTM detection
  • --skip-ptm: Skip PTM detection (faster, use if only instrument/tolerance info needed)

3.5 Keep downloaded files

  • --keep-files: Retain downloaded raw and converted mzML files after analysis
  • By default, files are cleaned up to save disk space

Step 4: Run techsdrf

4.1 Present the full command

Assemble the command with all chosen options and present it to the user:

techsdrf refine \
  -p PXD###### \
  -s input.sdrf.tsv \
  -n 5 \
  -o refined.sdrf.tsv \
  -v

4.2 Explain what happens

The techsdrf pipeline runs these stages:

  1. Download — Fetches N raw files from PRIDE (or reads local files)
  2. Convert — Converts raw files to mzML using ThermoRawFileParser or msconvert
  3. Analyze — Parses mzML with pyopenms to extract instrument metadata, scan characteristics, and spectra
  4. Detect — Determines instrument model, acquisition mode, fragmentation, tolerances, labels, and PTMs
  5. Compare — Compares detected parameters against declared values in the SDRF
  6. Refine — Produces a refined SDRF with corrected/added technical metadata

4.3 The user runs the command

The user executes the command in their terminal. techsdrf produces:

  • Console output with detected parameters and comparison results
  • A refined SDRF file (if -o was specified)

Step 5: Interpret Results

Parse the techsdrf output and explain each detected parameter:

5.1 Instrument detection

Detected Field Source Example
Instrument model Raw file header Q Exactive HF-X
Serial number Raw file header Exactive Series slot #1234
Mass analyzer Scan metadata orbitrap

What to check: Does the detected model match what's in the SDRF? PRIDE often has generic names (e.g., "Q Exactive") while raw files have the specific model (e.g., "Q Exactive HF-X").

5.2 Acquisition mode

Detected Field Method Example
DDA vs DIA Spectrum analysis Data-Dependent Acquisition

What to check: Is the acquisition method in the SDRF correct? Misclassification of DDA as DIA (or vice versa) affects which analysis pipelines can be used.

5.3 Fragmentation

Detected Field Source Example
Dissociation method MS2 scan metadata HCD
Collision energy Scan parameters 28 NCE

What to check: Does the SDRF declare the correct fragmentation method? Multiple methods may be used (e.g., HCD + EThcD for different scans).

5.4 Mass tolerances

Detected Field Method Example
Precursor tolerance Gaussian fitting (RunAssessor) + param-medic 10 ppm
Fragment tolerance Gaussian fitting (RunAssessor) + param-medic 20 ppm

techsdrf uses two independent methods to estimate tolerances. Both are reported for comparison. What to check: Do the detected tolerances match the SDRF? Papers often round tolerances or report search engine settings instead of instrument-level tolerances.

5.5 Labels

Detected Field Method Example
TMT/iTRAQ Reporter ion detection (100-140 m/z) TMT10plex
SILAC Known mass shift patterns not detected

What to check: Does the labeling in the SDRF match what's actually in the spectra? This catches cases where TMT is declared but the data is actually label-free (or vice versa).

5.6 PTM detection (three tiers)

Tier Method Example
1. Reporter/diagnostic ions Known ion signatures (100-140 m/z) TMT reporter ions
2. Diagnostic ions + neutral losses Characteristic fragment patterns Phospho (H3PO4 loss at 97.98 Da)
3. Open mass shifts Poisson-tested mass deltas Oxidation (+15.995 Da)

What to check: Are all detected PTMs declared as comment[modification parameters] in the SDRF? Are there declared modifications that were NOT detected (may indicate incorrect annotation)?

5.7 Comparison summary

techsdrf reports each parameter with a status:

  • MATCH — SDRF value matches detected value
  • MISMATCH — SDRF value differs from detected value (flag for correction)
  • MISSING_SDRF — Detected in raw data but not declared in SDRF (flag for addition)
  • IMPROVED — Detected value is more specific than SDRF value (e.g., "Q Exactive" → "Q Exactive HF")

List all MISMATCH and MISSING_SDRF items with recommended fixes.

Step 6: Apply Refinements

6.1 If techsdrf produced a refined SDRF file

Present the differences between original and refined SDRF:

CHANGED: comment[instrument] — "Q Exactive" → "Q Exactive HF-X" (AC=MS:1002877;NT=Q Exactive HF-X)
CHANGED: comment[precursor mass tolerance] — "20 ppm" → "10 ppm"
ADDED:   comment[dissociation method] — "HCD" (was empty)
CHANGED: comment[modification parameters] col 3 — REMOVED Phospho (not detected in spectra)

6.2 Let user approve/reject each change

For each changed column:

  • Show the old value and the new value
  • Explain WHY techsdrf made this change (what evidence from raw data)
  • Let the user approve or reject the individual change

6.3 If the user is working on an in-memory SDRF from /sdrf:annotate

Instead of a file diff, show the corrected values to paste into the SDRF:

Update these technical columns based on raw file analysis:
  comment[instrument]: AC=MS:1002877;NT=Q Exactive HF-X
  comment[precursor mass tolerance]: 10 ppm
  comment[fragment mass tolerance]: 20 ppm
  comment[dissociation method]: NT=HCD;AC=MS:1000422

Step 7: Post-Refinement

7.1 Update spec to latest version

Before validating, ensure the SDRF specification and templates are up to date:

git submodule update --remote --recursive

This pulls the latest column definitions (spec/sdrf-proteomics/TERMS.tsv) and template versions (spec/sdrf-proteomics/sdrf-templates/templates.yaml). Validation against outdated templates may miss new required columns or accept deprecated formats.

7.2 Validate with sdrf-pipelines

Always run programmatic validation on the refined SDRF before presenting results to the user:

parse_sdrf validate-sdrf \
  --sdrf_file refined.sdrf.tsv \
  --template <template1> \
  --template <template2>

Detect templates from comment[sdrf template] columns in the SDRF. If parse_sdrf is not installed, tell the user: pip install sdrf-pipelines

If validation finds errors introduced by the refinement, fix them and re-run until validation passes (or only warnings remain).

7.3 Contribute

If this is a ProteomeXchange dataset:

Your refined SDRF has verified technical metadata from actual raw file analysis.
Run /sdrf:contribute to submit this annotation to the community repository.

What techsdrf Detects

Parameter Detection Method
Instrument model + serial Raw file header parsing
DDA vs DIA Spectrum analysis (isolation window patterns)
Fragmentation method (HCD/CID/ETD/EThcD) MS2 scan metadata
Precursor mass tolerance Gaussian fitting (RunAssessor) + param-medic dual estimation
Fragment mass tolerance Gaussian fitting (RunAssessor) + param-medic dual estimation
Labels (TMT/iTRAQ) Reporter ion detection in 100-140 m/z range
PTMs (phospho, glyco, etc.) Diagnostic ions + neutral losses + open mass shifts
Variable modifications Open mass-shift analysis with Poisson statistical testing

Important Rules

  • techsdrf requires actual raw MS files — it cannot refine from metadata alone
  • For Thermo .raw files, ThermoRawFileParser must be installed
  • Detected parameters are empirical — they reflect what's IN the raw data, not what was intended
  • If techsdrf detects a PTM not declared in the SDRF, confirm with the user before adding it (could be a biological signal or an artifact)
  • If techsdrf does NOT detect a declared PTM, it may be at low abundance — do not automatically remove without user confirmation
  • Always verify instrument ontology terms via OLS after techsdrf detection:
    mcp OLS → searchClasses(query="Q Exactive HF-X", ontologyId="ms")
    